Discontinue using this mask if the child has any adverse reaction to
the use of the mask, and consult your physician or sleep therapist.
•
At a fixed rate of supplemental oxygen flow, the inhaled oxygen
concentration varies, depending on the pressure settings, patient
breathing pattern, mask, point of application and leak rate. This
warning applies to most type of CPAP or bilevel devices.
•
As with all masks, some rebreathing may occur at low CPAP pressures.
•
Refer to your device manual for details on settings and operational
information.
Technical specifications
Pressure–flow
curve
70
60
50
40
30
20
10
0
2
4
Physical dead
space
Therapy pressure
Compliance
Resistance
Environmental
conditions
The mask contains passive venting to protect against
rebreathing. As a result of manufacturing variations,
the vent flow rate may vary.
6
8
10
12
Mask pressure (cm H
101.7 mL
3 to 20 cm H
Mask compliance between 5 cm H
is less than 1.57 mL/cm H
Drop in pressure measured (nominal)
at 50 L/min: 0.8 cm H
at 100 L/min: 3.1 cm H
Operating temperature: +41°F to +104°F
(+5°C to +40°C)
Operating humidity: 15% to 95% RH non-condensing.
Storage and transport: -4°F to +140°F
( -20°C to +60°C)
Storage and transport humidity: up to 95% RH
non-condensing.
14
16
18
O)
2
O
2
O
2
O
2
O
2
17
Pressure
Flow
)
(cm H
O
(L/min)
2
3
18
7
28
12
36
16
43
20
20
49
O and 15 cm H
2
2
O