Electrostatic dis-
±6 kV contact
charge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for Input/
output lines
Surge
±1 kV line to line
IEC 61000-4-5
±2 kV line to earth
38
±6 kV
Floors should be wood,
contact
±8 kV air
are covered with synthetic mate-
rial, the relative humidity should be
at least 30%.
±2kV for power
Mains power quality should be that
supply lines
of a typical commercial or hospital
±1kV for interco-
environment.
nnecting cable
±1 kV line to line
Mains power quality should be that
of a typical commercial or hospital
environment.