(<10 cfu/ml) can be used for all rinsing steps. (For example, pure water that is
in accordance with the European Pharmacopoeia or the United States Pharma-
copoeia).
f) After cleaning, the chemical residue should be less than 10mg / L.
h) Regularly repair and inspect the disinfector.
6.5 Drying
If your cleaning and disinfection process does not have an automatic drying
function, dry it after cleaning and disinfection.
Methods:
compressed air (maximum pressure 3 bar). Until no liquid is sprayed onto the
white paper (white cloth), the product drying is completed.
2. It can also be dried directly in a medical drying cabinet (or oven). The recom-
mended drying temperature is 80C-120C and the time should be 15-40 minu-
tes.
Note:
a) The drying of product must be performed in a clean place.
b) The drying temperature should not exceed 138 °C;
c) The equipment used should be inspected and maintained regularly.
6.6 Inspection and maintenance
1. Check the product. If there is still visible stain on the product after cleaning/
disinfection, the entire cleaning/disinfection process must be repeated.
2. Check the product. If it is obviously damaged, smashed, detached, corroded
or ruptured, it must be scrapped and not allowed to continue to be used.
30