section in the next section "Disinfection".
Note:
a) The cleaning agent does not have to be pure water. It can be
distilled water, deionized water or multi-enzyme. But please ensure
that the selected cleaning agent is compatible with the product.
b) In washing stage, the water temperature should not exceed 45 °C,
c) After cleaning, the chemical residue should be less than 10mg / L.
6.4 Disinfection
Disinfection must be performed no later than 2 hours after the cleaning phase.
Automated disinfection is preferred if conditions permit.
Automated disinfection-Washer-disinfector
with EN ISO 15883.
·Use high temperature disinfection function. The temperature does not exceed
134
° C, and the disinfection under the temperature cannot exceed 20 minutes.
·The disinfection cycle is in accordance with the disinfection cycle in EN ISO
15883.
Cleaning and disinfecting steps by using Washer-disinfector:
1. Carefully place the product into the disinfection basket. Fixation of product
is needed only when the product is removable in the device. The products are
not allowed to contact each other.
2. Use a suitable rinsing adaptor, and connect the internal water lines to the
rinsing connection of the washer-disinfector.
28