Note:
a) Only products that have been effectively cleaned and disinfected are
allowed to be sterilized;
b) Before using the sterilizer for sterilization, read the Instruction Ma
nual provided by the equipment manufacturer and follow the instruc
tions.
c) Do not use hot air sterilization and radiation sterilization as this may
result in damage to the product;
d) Please use the recommended sterilization procedures for
sterilization. It is not recommended to sterilize with other
sterilization procedures such as ethylene oxide, formaldehyde and
low temperature plasma sterilization. The manufacturer assumes no
responsibility for the procedures that have not been recommended.
If you use the sterilization procedures that have not been recommen
ded, please adhere to related effective standards and verify the suita
bility and effectiveness.
*Fractional pre-vacuum procedure = steam sterilization with repetitive pre- va-
cuum. The procedure used here is to perform steam sterilization through three
pre- vacuums.
6.9 Storage
1. Store in a clean, dry, ventilated, non-corrosive atmosphere with a relative
humidity of 10% to 93%, an atmospheric pressure of 70 kPa to 106KPa, and a
temperature of -20 °C to +55 °C;
2. After sterilization, the product should be packaged in a medical sterilization
bag and stored in a special storage cabinet. The storage time should not ex-
ceed 7 days. If it is exceeded, it should be reprocessed before use.