Problems
The length of the root canal
cannot be measured.
No sound of alarm.
The charging LED
indicator goes out.
22
Possible causes
1. If the measuring wire is
connected correctly?
2. If the measuring wire is
broken?
If the volume is set at "mute"?
1.The adapter is not connec-
ted well.
2.Have used faulty adapter
with excessive output.
3.The battery is not installed
well.
4.The battery has been
damaged.
Solutions
plugged
to check if the measuring wire
is broken.
Adjust the sound level.
1.Reconnect the adapter.
2.Change the adapter, must
use the original
adapter.
3.Reinsert the battery and
then reconnect
the adapter.
4.Change the battery and then
reconnect
the adapter.