3. Check the product. If the accessories are found to be damaged, please re-
place it before use. And the new accessories for replacement must be cleaned,
disinfected and dried.
service life (number of times), please replace it in time.
6.7 Packaging
Install the disinfected and dried product and quickly package it in a medical
sterilization bag (or special holder, sterile box).
Note:
a) The package used conforms to ISO 11607;
permeability;
c) The packaging environment and related tools must be cleaned regu
larly to ensure cleanliness and prevent the introduction of contami
nants;
d) Avoid contact with parts of different metals when packaging.
6.8 Sterilization
Use only the following steam sterilization procedures (fractional pre-vacuum
procedure*) for sterilization, and other sterilization procedures are prohibited:
285 to comply with EN ISO 17665;
2. The highest sterilization temperature is 138 ° C;
3. The sterilization time is at least 4 minutes at a temperature of
132 ° C / 134 ° C and a pressure of 2.0 bar ~ 2.3 bars.
4. Allow a maximum sterilization time of 20 minutes at 134 °C.