4 Preparation
24
4.3
Before each patient
1. Purge the water and air channels for 20-30 seconds.
Before each patient
Preparation _ product name
2. Connect BA ULTIMATE.
3. Insert the burr instrument [ → 29].
4. Use a sufficient amount of cooling water (> 50 ml/min).
5. Use filtered water only (< 50µm).
6. Check the nozzles for blockages and lime deposits, for example, and
clean the nozzles if necessary.
CAUTION! Insufficient cooling leads to overheating of the
Overheating at the preparation site
preparation site and damage to the tooth substance. Ensure that the
water content is > 50ml/min.