Before Each Patient - B.A. International BA45 Instrucciones De Uso

Ocultar thumbs Ver también para BA45:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

6453505.book Seite 24 Donnerstag, 21. April 2016 4:59 16
4 Preparation
24
4.3

Before each patient

1. Purge the water and air channels for 20-30 seconds.
Before each patient
Preparation _ product name
2. Connect BA ULTIMATE.
3. Insert the burr instrument [ → 29].
4. Use a sufficient amount of cooling water (> 50 ml/min).
5. Use filtered water only (< 50µm).
6. Check the nozzles for blockages and lime deposits, for example, and
clean the nozzles if necessary.
CAUTION! Insufficient cooling leads to overheating of the
Overheating at the preparation site
preparation site and damage to the tooth substance. Ensure that the
water content is > 50ml/min.
Operating Instructions BA ULTIMATE Kit
64 53 505 D3551.201.04.02.02
Tabla de contenido
loading

Este manual también es adecuado para:

Ba175Ba605Ba606Ba695fbBa695fm

Tabla de contenido