Maintenance
Take into account the content of
Confirm periodically that weighing value are correct.
Adjust the scale if necessary.
"1.1. Detailed
Refer to
adjustment of zero point".
Troubleshooting checklist and solutions
Problem
The power does not turn on.
Nothing is displayed.
Zero is not displayed when the
display is turned on.
The display does not respond.
Counting mode cannot be used.
Error codes
Error codes
Unstable weighing value
"Zero display" and "sensitivity adjustment" cannot be performed.
Input error
The value input for unit weight or tare value is out of range.
Input a value within the range.
The memory (circuit) has malfunctioned.
The voltage sensor has malfunctioned.
Weighing sensor error
Sensitivity adjustment error
The sensitivity adjustment has been stopped because the sensitivity
adjustment weight is too heavy or too light. Use a proper sensitivity
adjustment weight and adjust the scale.
English (EN)
"2.1. Precautions when installing the
manual" for "sensitivity adjustment" and "sensitivity
Confirm that nothing is touching to the weighing pan.
Avoid breezes and vibration.
Perform "sensitivity adjustment of zero point".
Press the RESET key to return to weighing display.
Confirm that cable between display unit and weighing unit is connected
correctly.
The weighing sensor has malfunctioned.
Check items and solutions
Confirm that the AC adapter is connected correctly.
Confirm that nothing is touching to the weighing
pan.
Remove anything on the weighing pan.
Perform sensitivity adjustment of zero point.
Turn off the display and then turn it back on.
Confirm that unit weight has been entered.
Refer to
"4. Counting
Descriptions and solutions
11
scale".
mode".