Error code
5. Error Codes
EC, E11 Stability error
The balance cannot stabilize due to an environmental problem.
Check around the pan. Prevent vibration, drafts, temperature
changes, static electricity and magnetic fields, from influencing the
balance. Refer to "3-2 During Use" in the instruction manual.
To return to the weighing mode, press the
Out of range error
The entered value exceeds the setting range.
Enter a correct value.
EC, E17 Internal mass error (FZ-i / FZ-i WP / FZ-GD series only)
The internal mass application mechanism does not work properly.
Perform the weighing operation from the beginning again.
EC, E20 Calibration weight error (Positive value)
The calibration weight is too heavy. Verify the calibration mass
value. Press the
EC, E21 Calibration weight error (Negative value)
The calibration weight is too light. Confirm the calibration mass
value. Press the
Overload error
A sample beyond the balance weighing capacity has been placed
on the pan. Remove the sample from the pan.
Weighing pan error
The mass value is too light. Confirm that the weighing pan is
properly installed and calibrate the balance.
Sample mass error
The balance cannot store the sample for the counting mode or for
the percent mode because it is too light. Use a larger sample.
Unit mass error
The sample unit mass for the counting mode is too light. Storing and
using it for counting will cause a counting error. Add samples to
reach the specified number and press the
the
the counting mode. But, for accurate counting, be sure to add
samples.
Clock battery error
The clock backup battery has been depleted. Press any key and set
the time and date. The clock and calendar function work normally as
long as the AC adapter is connected to the balance. If this error
appears frequently, contact the local A&D dealer.
Description
key to return to the weighing mode.
key to return to the weighing mode.
key without adding samples will shift the balance to
9
key.
key. Pressing