CAUTION
– In cleaning procedure the cleanser should be neutralization agent, never use the cleanser or disinfectant containing chlorine.
– Do not immerse the head of the medical device or the medical device to a disinfectant bath, and it is not permitted to clean the medical device in
the ultrasonic unit.
– Autoclaving can be operated in a steam sterilizer that corresponds with EN13060 B/ISO17665-1 Class B or S and is specifi ed for the dental turbines.
– Autoclaving maximum temperature does not exceed 138°(280°F)
– Remove the medical device from the steam sterilizer and store in a dry, dark and cool space protected from dust with minimum exposure to germs.
– Do not immerse the medical device in cold water to speed up the cooling process after removing from the steam sterilizer.
10. Storage
– Reprocessed medical device should be stored in a dry, dark and cool space protected from dust with minimum exposure to germs, corrosive gas
and far away from the heat source.
– Comply with the expiry date of the sterilized items.
+55°C
-40°C
38
80%
106 kPa
10%
min.
50 kPa