Nature and Frequency of Preventative Maintenance

Inspection and Care of the Toto Touch Control Unit
The Toto Touch control unit is designed to be reliable and long lasting, with few parts requiring
maintenance.
A routine service is required when the service light illuminates. Please refer to the Toto Service,
Maintenance and Repair Manual for further information. Supporting videos are also available: https://
bit.ly/TotoVideos
Clean air filters at least every 12 months. Inspections should be undertaken to note any visible signs
of damage.
In the event of product failure, please contact Frontier Therapeutics Limited for advice and
recommendations on service and repair.
Although there are no requirements for Portable Appliance Testing, guideline recommendations
indicate that extension leads and portable electrical equipment should be re-tested every six months.
Please refer to local guidance.
An electrical safety test in accordance with BS EN 62353:2014 Medical Electrical Equipment -
Recurrent Test and Test After Repair of Medical Electrical Equipment, should be carried out after
maintenance, inspection, servicing and repair, and before the control unit is sent out to, or returned
to a user.

Cleaning and Disinfection

All components of the system can be cleaned with detergent and water to remove any visible
contamination and chemically disinfected with chlorine.
The following represents guidance on the correct cleaning and decontamination process but does
not replace local policy and guidelines.
Toto Lateral Turning System Platform Cover
Keeping a cleaning record is recommended as the Toto system is suitable for re-use.
Light and Heavy Soiling
Decontaminate with a 0.1% (1,000ppm) - 1% (10,000ppm) Chlorine solution.
Rinse with clean water and wipe down with a single-use, non-abrasive cloth and thoroughly dry.
Do not apply a 1% chlorine-based solution for more than two minutes, doing so may cause long
term damage.
Prolonged use of alcohol-based cleaning agents may reduce the life of the product.
If used thoroughly rinse with clean water and allow to dry before use.