Dear Customers
Congratulations on purchasing an A&D blood pressure monitor. This
device is designed for ease of use and accuracy.
We recommend that you read through this manual carefully before
using the device for the first time.
Introduction
Intended user: The device is designed to be used by medical workers.
The device is designed for use on adults and children of age 3 years and
older.
Environment for use: The device is designed to be used in medical
facilities.
This device is designed to measure blood pressure and pulse rate of
people for diagnosis.
Operating principle: This device is an oscillometric sphygmomanometer.
When the cuff pressure is pressurized above systolic blood pressure and
then is gradually depressurized, the device can detect the pulse
synchronized to heart beat included in the cuff pressure. This pulsation
starts small at the beginning, becomes large as the pressure decreases,
eventually shows the maximum amplitude, then decreases again, and
ultimately creates a mountain-shaped pattern. This oscillometric-type
sphygmomanometer analyzes the signal information of this pulsation
with a microcomputer and determines systolic blood pressure and
diastolic blood pressure values.
This device complies with part 15 of the FCC rules and contains the
FCC ID VPYLBZY.
English 3