7
TROUBLESHOOTING
7.1
Pump will not activate:
Check pump output terminals for loose screws and disconnected wires.
Check for proper voltage across motor windings.
Check for obstruction in pump head.
Check the program time and quantity and also clock time.
7.2
Pump runs too slowly:
Check roller block for binding.
Check for lubrication on squeeze tube.
Pump was setup at the lower speed.
7.3
Loss of pump prime:
Check pickup line for any holes or air leaks.
Check squeeze tubing in pump for any cracks or pin holes.
Check tubing for deterioration.
8
QUICK REFERENCE
Switch ON
At the first time the system is switched ON, it will ask the following settings:
1. Language (you can choose EN, IT, FR, DE, ES).
2. Clock Setup
3. Unit (you can choose ml or oz).
Priming
After the hydraulic installation, keep pressed UP key for at least 2 seconds until the pump start to run at the max
speed, press again to stop the pump.
Calibration
The system is already calibrated for 15 l/h (8 oz/min) with water and no back pressure, because of that calibration
is recommended in any case.
To do calibration keep pressed Set key for 3 sec than UP and DOWN until you reach "Pump Calibration" window.
Programming Menu
P01
-------
P01
-------
P01
-------
P01
-------
P01
Durat.
H E A TS A V R
on
12.00am
Q
0ml
on
12.00am
Q
0ml
on
12.00am
Q
0ml
on
12.00am
Q
0ml
Speed 100%
00:00:00
C od e 0000137248
st
1
Number of program - You can set 48 different
programs. Select the program number, you want start to
setup by pressing UP and DOWN keys. To confirm press
Enter.
nd
2
Set startup time - The dosing start time can be set by
using UP and DOWN keys, to confirm press Enter.
rd
3
Weekly days – Use UP and DOWN to choose which
weekly day or group of days you want the program running.
Press Enter to confirm.
th
4
Quantity - Use UP and DOWN keys to set the desired
quantity, to confirm press Enter
The maximum quantity which can be set is 18 l or 600 oz.
*remember to setup the measure unit of the system.
th
5
Pump speed - Use UP and DOWN keys to choose the
pump speed expressed as a percentage (from 1 to 100%)
which automatically fixes the duration of the dosing itself.
The bottom right part of the display shows the time
necessary for dosing at that given speed.
To confirm press Enter.
rev . 1 . 0
8