Note: Refer to the PACE User Manual, Reference and Specification for details of other system
components.
*
High pressure gas exhaust - depending on pressure range.
**
Optional vacuum system kit, allows the -ve port gas to be directly discharged to atmosphere, by-
passing the vacuum pump.
Optimum controller transient response and minimum time to set-point may be degraded if either the
pneumatic supply or vacuum system has restricted flow. Installing a reservoir volume, which has larger
capacity than the load volume, located in close proximity to the controller supply ports can improve the
controller response.
Optional negative gauge pressure generator kit, allows the -ve port to directly discharge to
atmosphere, by-passing the negative gauge pressure generator.
To protect the control module, for ranges above 100 bar (1450 psi), from over-pressure a suitable
protection device (such as a relief valve or bursting disc) must be fitted to limit the applied supply
pressure to below the Maximum Working Pressure (MWP).
Optional differential connection kit.
3. Operation
After the power-up sequence, the instrument shows the default display on the touch screen. The
touch screen divides into a number of mimic keys.
3.1 PACE5000 Single Chanel Display
Refer to Figure D1 and the key below:
1.
Status
2.
Switches between measure and control modes
3.
Enter new set-point value
4.
Controller set-up menu
5.
Measure mode set-up menu
3.2 PACE6000 Single Chanel Display
Refer to Figure D2 and the key below:
1.
Status
2.
Switches between measure and control modes
3.
Enter new set-point value
4.
Controller set-up menu
5.
Measure mode set-up menu
Note: The PACE6000 can be set to show a single channel display.
3.3 Setpoint Functional Description
Refer to Figure D3 and the key below:
1.
Back space (deletes last entered character)
2.
Switches positive/negative value
3.
Enters decimal point
4.
Escape - exits this menu
5.
Selects new digit for set-point value
6.
Accepts (enters) new complete set-point value
Copyright 2008 Baker Hughes Company.
8 | PACE5000/6000 Safety Instructions–English