Message
Control solution
test result is out
of range
Never try to disassemble the meter under any
circumstances. If you encounter any error messages
not listed above or if you have followed the actions
recommended but the problem remains unsolved,
please call Customer Support 24/7 at 1-866-373-2824.
What it means
Error in performing
the test.
Did not shake the
control solution vial
very well.
Expired or conta-
minated control
solution.
Control solution that
is too warm or too
cold.
Control solution is
stored outside of
the specified
temperature range,
which is from
35°F to 86°F(2°C to
30°C).
Test strip
deterioration.
Defective meter.
31
What to do
Read the instructions
thoroughly and repeat
the test again.
Shake the control
solution vigorously and
repeat the test again.
Check the expiration
date or the discard date
of the control solution.
Control solution, meter,
and test strips should
come to room tempe-
rature 68-77°F(20-
25°C) before testing.
Use a new bottle of
control solution that
has been stored at
specified temperature
range.
Repeat the test with
a new test strip.
Please call Customer
Support for help.