Quality Control

PetTest Level 1, Level 2 and Level 3 Control Solutions are
used to verify that the meter and test strips are working
properly together as a system. It is very important that you do
this simple check routinely to make sure results are accurate.

When to check your system with control solutions

Before using the meter for the first time.
When you begin using a new vial of test strips.
Whenever you suspect that the meter or test strips are not
working properly.
After cleaning your meter.
At least once a week.
To check your technique.
When you think results might be wrong.
When your test strips are exposed to extreme environmental
conditions (below 39.2°F(4
above 85% R.H. or below 15% R.H.).
If you drop the meter.
If you leave the strip vial cap opened.