4. Hold the meter to touch the drop until the meter beeps and
begins to count down. To avoid contaminating the control
solution with the content of the test strip, DO NOT directly
apply control solution into a strip.
5. After 5 seconds, the control solution test result appears.
Compare the result with the range printed on the test strip
vial. The result should fall within this range. Remove the
used strip, the meter will automatically turn off.
If the result is not in the acceptable range, it may be
caused by:
Performing the test improperly.
Not shaking the control solution vial well enough.
Using expired or contaminated control solution or test strips.
Control solution is stored at environment out of the specified
temperature range, which is from 35°F to 86°F(2°C -30°C).
Test strips are stored at environment out of the specified
temperature range or humidity range, which are from 39.2°F
to 104°F(4°C -40°C), 15~85% R.H (non-condensing).
Using the first drop of control solution.
Using a damaged meter.
If you continue to get control solution test results that fall
outside the range, the system may be not working properly.
DO NOT use the system to test blood glucose.
Contact Customer Support 24/7 at 1-866-373-2824 for help.