Limitations - Advocate PetTest BMB-EA099X Manual Del Usuario

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Limitations

A. The test strips are to be used with fresh, capillary whole
blood samples.
B. DO NOT perform glucose tests on your pets when your pets
are sick or in poor body conditions.
C. DO NOT use to test newborn animals (neonates).
D. Not to be used for determination of a diagnosis or screening
of diabetes.
E. Extreme humidity may affect the results. A relative humidity
greater than 85% or lower than 15% may cause incorrect
results.
F. The system should be used at a temperature between
50°F-104°F (10°C - 40°C). Outside this range, the system
will cease operation. Bring system within operating
temperature range for 20 minutes before using.
G. DO NOT reuse the test strips. The test strips are for single
use only.
H. Red blood cell count (Hematocrit): Accurate results are
provided for hematocrit ranges between 20% and 60%;
hematocrit levels below 20% may cause higher results;
hematocrit levels above 60% may cause lower results. If
you do not know hematocrit level of your pets, please
consult with your veterinarian.