Maintenance

1. Cleaning Procedure

1-1. Take disinfectant CaviWipesXL towelette
from the container.
1-2. Wipe down the entire meter and lancing
device, including the clear cap, until the
debris or the bioburden is removed. After
air drying, go to Step 2: Disinfection
Procedure.
1-3. Discard the used disinfectant wipe in trash.

2. Disinfection Procedure

Perform the cleaning procedure above prior to performing
the disinfection procedure.
2-1. Take a new disinfecting wipe or towelette
from the container.
2-2. Wipe down the entire meter enough to
thoroughly wet all surfaces (3 passes
horizontal and 3 passes vertical as
shown below) to complete the
disinfection procedure. Allow the
devices to remain wet for a 2-minute
contact time. Perform a similar
procedure on the entire surface of the
lancing device, including the cap.
Do disinfection at least once per week.