During the countdown :
✓
•
Pressing
brings the user back to the test.
•
Pressing
cancels the test and reinitializes the machine.
NB: If the technician has not pressed
(i.e. he has not resumed the test), the test is canceled and has to be restarted at the
beginning.
✓
Pressing of
cancels the test and brings the machine back to the intitialization step.
7.12 Registration of protocol data
At the end of a test process, the Neptune instrument beeps and the touchscreen displays :
The USB stick used has to be on FAT (File Allocation Table) format.
To safeguard the data of the protocol which has just finished, press on USB. The screen
displays the steps that are undertaken automatically:
•
the screen asks the user to plug in the USB Stick (« Insert USB »)
•
The system checks the remaining space on the USB stick (« Do not remove » is
displayed)
Note: the analysis of the USB stick can take a certain time. Please wait at least 2 minutes
before removing the USB stick. The touchscreen displays that the USB stick can be removed.
•
Writing of the data (« Writing USB »)
•
Removal of the USB stick (« Remove USB »)
After recording, the program proposes to make an additional copy.
It is strongly advised to safeguard the data of the test. However, if the user does not wish to
carry out this recording, the pressing of
requests a confirmation of the not-recording.
Neptune Instrument
✓
before the end of the countdown of the 30 seconds
allows to skip this step. The programs however
22