Make sure that the unit is working properly before it is put into service . The following list will ensure that each part of
the unit is checked .
1 .
Fully depress the brake pedal of the stretcher to set the four wheel brakes and verify that all four casters are
locked .
2 .
Raise and lower the hydraulic lift system .
3 .
Raise the unit completely and activate the Trendelenburg function . Ensure that the head end lowers to the lowest
position .
4 .
Raise the unit completely and activate the reverse Trendelenburg function . Ensure that the foot end lowers to the
lowest position .
5 .
Activate the fifth wheel to ensure that it is operating properly .
6 .
Ensure that the siderails raise and lower smoothly and lock securely when fully raised .
7 .
Raise and lower the fowler (head end) .
CAUTION
Do not modify this stretcher . Modifying the unit can cause unpredictable operation resulting in injury to the patient or
operator . Modifying the unit will also void its warranty .