OBJ_DOKU-184654-006.fm Seite 22 Montag, 1. Februar 2021 11:28 11
1.4
Qualification of user
For the intended use, the product may only be operated by med-
ical professionals who have been trained in its use on the basis
of the notes on use.
1.5
Performance characteristics
Performance characteristics relating to the intended use are:
• Transporting HF current from the electrosurgical unit to the
instrument.
• Transporting the argon gas from the APC unit to the instru-
ment.
2
Maximum electrical capacity
The maximum electrical capacity of this product is:
• 4300 V
p
3
Safety instructions
WARNING!
Check the product for external damage before each use.
Erbe Elektromedizin recommends performing an electrical con-
tinuity test before every use of this product.
Note: If the cable conductor is broken or the electrical continuity
of the cable is interrupted in any other manner it can cause arc-
ing in the patient return line or in the active circuit; this can lead
to the patient suffering burns or the outbreak of a fire.
If damaged, do not use this product!
22