OBJ_DOKU-193153-006.fm Seite 23 Mittwoch, 25. März 2020 4:12 16
5.2
User requirements
This product may be used only by trained medical staff, who
have been shown how to use it according to the instructions for
use.
Only use the product if you are familiar with its handling in con-
junction with the respective APC and electrosurgical unit!
5.3
General information about protection of the user,
patient, and product
Do not use in the presence of combustible or explosive materi-
als!
Never lay this product on the patient or in his/her direct vicinity.
Protect this product from any form of mechanical damage! Do
not throw! Do not use force! If there is a cable, and unless stated
otherwise, do not kink it or wrap it around the accompanying
product!
Do not kink the tube or wrap it round the product to which it be-
longs!
If damaged, do not use this product!
5.4
Risk of fire in the gastrointestinal tract
Endogenous gases (e.g. methane) can ignite through the HF flow
if this product is used in the gastrointestinal track. It must there-
fore be ensured, prior to use of the product that no endogenous
gases are present in the gastrointestinal track.
23