Expected Lifetime
The expected lifetime of the Plié 3 MPC Knee is defined by the warranty period and contingent on adherence to the service
plan.
3. INTENDED USE/INDICATIONS
The Plié 3 MPC Knee is appropriate for users who would benefit from the safety inherent in the stability of a microprocessor-
controlled knee. This medical device is supplied to healthcare professionals (prosthetists) who will train the patient in its use. The
prescription is made by a doctor together with the prosthetist, who assess the patient's ability to use it.
This device is for SINGLE PATIENT use. It should not be reused on another patient.
4. CLINICAL BENEFITS
The Plié 3 MPC Knee is appropriate for users who would benefit from the safety inherent in the stability of a
microprocessor-controlled knee. These users should also have the ability or have the potential to:
•
negotiate obstacles in the community or workplace
•
exert sufficient hip joint or pelvic voluntary muscle control
•
ambulate with variable cadence
•
descend stairs and ramps
5. ACCESSORIES AND COMPATIBILITIES
The device incorporates a male pyramid link that allows it to be used with female pyramid connectors for the Pyramid Top version and a
M36 x 1.5 thread for the Threaded Top version.
6. ASSEMBLY AND PATIENT FITTING
A. Battery and Battery Cap
•
Follow the Plié 3 Owner's Guide on battery handling and charging.
•
If power is lost from the Plié 3 knee, it will default to the Stance Flexion resistance setting and will not release into swing.
•
Insert a charged battery when the Plié 3 MPC Knee low battery indicator is visible (symbolized by red blinking light visible
on the left side of the knee viewed from posterior) prior to complete power loss to prevent potential personal injury.
•
Use ONLY compatible Plié 3 MPC Knee batteries and charger.
•
Store spare battery in the battery bag to avoid the risk of battery short circuit.
•
Use a lint free cloth with or without isopropyl alcohol to wipe the battery compartment area and gasket on the bottom
side of the cap (see pictures below). These areas need to be debris and lint free.
Battery
Compartment
Area
The Plié® 3 MPC Knee is intended for use as a component in a prosthetic leg for individuals with
lower-limb loss including:
•
transfemoral amputees
•
knee disarticulation amputees
•
hip disarticulation amputees
•
individuals with congenital lower-limb abnormalities
It is recommended specifically for patients with medium activity level (2.5) to high activity patients
(4):Maximum weight (load carrying included):
•
for moderate activity: 125 kg / 275 lbs.
•
for high activity: 100 kg / 220 lbs.
Gasket on Bottom
Side of Cap