Note: When date and time is set for first time, In-Service date is also set and is used for calculating
initial calibration interval (see "Setting Calibration Interval" in Advanced Configuration section).
Note: If wrench is powered up in torque only measurement mode, angle is not zeroed until mode is
changed to angle measurement mode, at which time torque and angle zeroing begins automatically
after 2 seconds. Wrench should be placed on a stable surface with no torque applied.
Note: Pressing ENTER button while angle is zeroing will abort zeroing function to allow user to
select another measurement mode.
TORQUE MODE
1. Set Target.
Use UP/DOWN buttons to change TORQUE target value.
2. Select Units of Measure.
Repeatedly press UNITS button while on target TORQUE screen until desired units are displayed.
3. Apply TORQUE.
Grasp center of handle and slowly apply torque to fastener until progress lights display green and a
½ second audible alert and handle vibration alerts you to stop.
4. Release TORQUE.
Note peak TORQUE reading flashing on LCD display for 5 seconds. Pressing BACKLIGHT button
while peak torque is flashing will continue to display value until button is released. Momentarily
pressing UP/DOWN, ENTER or UNITS button will immediately return to target TORQUE screen.
Reapplying TORQUE will immediately start another TORQUE measurement cycle.
5. Recall Peak TORQUE Reading.
To recall last peak TORQUE measurement, press and hold BACKLIGHT button for approximately 3
seconds. Peak TORQUE will flash for 5 seconds.