Setting Date And Time; Calibration; User Calibration - VWR MCP 100 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 59
E 132
Store
Date
2
3
Time
Cal
Mode
SET
DATE
YYYY
(2 sec.)
Mode
Zero
Test
!
Store
Date
Cal
4
Time
Cal
cAL
CAL
CAL
METHOD
Zero
Test
METHOD
0.0.0
CAL
Zero
Test
METHOD
VWR ver 1 23.01.2018
instrument switches off.
If the instrument is not able to communicate with the IRIM, a timeout
occurs after approx. 2 minutes. The error E 132 is displayed for approx.
4 seconds. Subsequently, the instrument switches to test mode (see
also IRIM manual).
705-1086
3
Setting date and time (24-hour-format)
2
After confirming the selection with the [MODE] key the value to be
edited will be shown for 2 sec.
The setting starts with the year (YYYY) followed by the actual value
to be edited. The same applies for month (mm), day (dd), hour (hh)
and minutes (mm). Set the minutes first in steps of 10, press the [!]
key to continue setting the minutes in steps of 1.
Increase the value by pressing the [MODE] key.
Decrease the value by pressing [ZERO/TEST] key.
Proceed to the next value to be edited by pressing [!] key.
After setting the minutes and pressing the [!] key the display will show
"IS SET" and the instrument returns to the measurement mode.

User calibration

Note:
user calibration (Display in calibration mode)
factory calibration (Display in calibration mode)
After confirming the selection with the [MODE] key the instrument will
show CAL/"Method".
Scroll through methods using the [MODE] key.
Fill a clean vial with the standard up to the 10 ml mark, screw the cap
on and place the vial in the sample chamber making sure that the
marks are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 8 seconds.
The display shows the following in alternating mode:
Perform calibration with a standard of known concentration (see
"Operation").
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
Infrared Data Transfer Modul IRIM
13
Tabla de contenido
loading