Troubleshooting
Pump does not start when starting a process despite required pressure re-
duction.
Cause
Overtemperature pro-
tection of the vacuum
system has been trig-
gered
Vacuum system does not pump
Cause
Connections or lines
are blocked.
External valve is closed
or filter is clogged.
62
Tab.23: Troubleshooting: Sufficient vacuum is not achieved despite
running pump
Troubleshooting
à
Disconnect the vacuum system from the electrical
mains.
à
Allow the pump to cool down.
à
Determine the cause of the overheating and rectify.
Tab.24: Troubleshooting: Pump does not start when starting a
process despite required pressure reduction
Troubleshooting
à
Check the connections and lines.
à
Remove the blockage.
à
Check external valves and filters.
Tab.25: Troubleshooting: Vacuum system does not pump
Translation of Originial Operating Instruction, KNF 325562-325564 06/21
Vacuum system SH820G / SR820G SH840G / SR840G