Vacuum system SH820G / SR820G SH840G / SR840G
Requirements
Material and
tools
Information on
the procedure
Translation of Originial Operating Instruction, KNF 325562-325564 06/21
9.3 Replace diaphragm, valve plates/seals and
O-rings
Necessary requirements
Vacuum sys-
- Vacuum system switched off and mains
tem
plug pulled out of socket
- Vacuum system cleaned and free of haz-
ardous materials
- Hoses removed from pneumatic inlet and
outlet
Quan-
Material
tity
1
TORX® T20 screwdriver with torque indicator
1
Open-end wrench, size 14
1
TORX ® screwdriver T25 with torque indicator (only
for SH840G and SR840G)
1
Spare parts set (see Chapter 10 Spare parts and ac-
cessories [} 57])
1
Pencil
Tab.16:
à
Always replace diaphragms, valve plates/seals, and O-
rings together to maintain the performance of the pump.
à
Replace the diaphragms and valve plates/seals of the in-
dividual pump heads one after the other.
à
As standard, only the elastomer parts of the pump are re-
placed during servicing. For servicing of the complete vac-
uum system, please refer to Chapter 9.4 Change O-rings
on the complete vacuum system (optional) [} 52].
Servicing
DE
EN
FR
ES
IT