Menu options
GB
Menu selections
Press the [MODE] key and hold.
Mode
Switch the unit on using the [ON/OFF] key.
On
Allow the 3 decimal points to be displayed before releasing the
Off
[MODE] key.
!
The [!] key allows for selection of the following menu points:
1
Store
Date
2
1
2
3
4
4
Time
Cal
The selected menu is indicated by an arrow in the display.
Mode
Confirm the selection with the [MODE] key.
The photometer shows the last 16 data sets in the following format
(automatically proceeds every 3 seconds until result is displayed):
Number
Year
Date
Time
Test
Result
Zero
Test
The [ZERO/TEST] key repeats the current data set.
Mode
The [MODE] key scrolls through all stored data sets.
!
Quit the menu by pressing [!] key.
After confirming the selection with the [MODE] key the value to be
Mode
edited will be shown for 2 sec.
SET
The setting starts with the year (YYYY) followed by the actual value
to be edited. Same applies for month (mm), day (dd), hour (hh) and
DATE
minutes (mm). Set the minutes first in steps of 10, press the [!] key to
YYYY
continue setting the minutes in steps of 1.
(2 sec.)
Increase the value by pressing the [MODE] key.
Mode
Decrease the value by pressing [ZERO/TEST] key.
Zero
Test
Proceed to the next value to be edited by pressing [!] key.
!
After setting the minutes and pressing the [!] key the display will show
"IS SET" and instrument returns to the measurement mode.
ATTENTION: If the battery is removed for more than one minute
the date and time menu starts automatically when the photometer is
switched on the next time.
34
All manuals and user guides at all-guides.com
recall stored data
setting the date and time
3
user calibration
Recall of stored data
1
n xx (xx: 16...1)
YYYY (i.e. 2008)
mm.dd (monthmonth:dayday)
hh:mm (hourhour:minuteminute)
Method
x,xx
3
Setting date and time (24-hour-format)
2
CheckitDirect_Cl-pH-CyA-TA_2 09/2010