Connection to an APC 2 with APC socket
IMPORTANT! The adapter, Art. No. 20132-227, must be plugged
into the APC receptacle!
1. Switch on the VIO electrosurgical unit and select the basic
program.
2. Insert the APC connector (1) into the adapter (2).
2
1
3. On the VIO electrosurgical unit, a standard setting is config-
ured for mode, power, effect, and gas flow. This standard
setting is the same for all APC probes. In order to achieve the
desired tissue effect, you may therefore have to adjust the
mode, power, effect, and/or gas flow to the APC probe being
used. The gas flow can only be changed within a certain
range. Mode, power, and effect can be changed as required.