5. Rinse the surfaces with sterile, deionized water (at least
1 minute).
6. Completely fill a disposable syringe with sterile, deionized
water and rinse the lumina (at least 5 times).
7. Repeat the preceding cleaning steps if the product or used
rinse water still contains visible contamination.
8. Immerse the product completely in a disinfectant bath, en-
suring that the product does not come into contact with oth-
er components in the bath.
9. Completely fill a disposable syringe with disinfecting solution
and rinse the lumina (at least 5 times).
10. Rinse the surfaces with sterile, deionized water (at least
1 minute).
11. Completely fill a disposable syringe with sterile, deionized
water and rinse the lumina (at least 5 times).
12. Remove the irrigation adapter.
13. Dry the product until there are no more fluid residues visible
(e.g. with filtered compressed air).
5.7
Cleaning and disinfection by machine
The washer/disinfector must have fundamentally certified ef-
fectiveness (e.g. CE marking according to DIN EN ISO 15883).
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have a pH value between 5.5
and 11.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
1. Place the product in a suitable rinsing basket carefully. In do-
ing so make sure the product is not touching other instru-
ments or parts of instruments.
2. Connect the product via the irrigation adapter to a rinsing
station.
26