CARE / MAINTENANCE / STORAGE
General:
Clean the device with a soft dry cloth.
Do not use abrasives or volatile cleaning agents.
Switch off the device when it is not in use.
Before storing the Promed INH-2.1 for a longer period, remove the batteries from the
battery compartment. Leaking batteries may damage the device.
Store the device and its accessories in the supplied storage bag in a cool, dry place. Please
ensure that neither the mouthpiece nor an inhalation mask is attached and that the protective
mask is always on the medicament holder.
Do not expose the device to direct sunlight and protect it against dirt and moisture.
For inspection and recalibration, the device does not have to be sent to the dealer or the
manufacturer.
If you detect other problems, please contact your dealer who sends in the device, if necessary.
Never try to correct an error on your own.
Note:
The device, particularly the medicament holder, must be cleaned with distilled water after each use.
Cleaning:
DE
A.
Removing inhalation solution residues.
Remove the medicament holder from the main device and empty the holder. Fill the holder with
a small amount of distilled water and re-attach it to the main device. Now switch on the device
in the cleaning mode by pressing and holding the Start/Stop button (F) until the operating
EN
mode LED (G) lights up blue. Switch off the device after 2 minutes and empty the medicament
holder as described previously.
B.
Cleaning the mesh membrane
To clean the mesh membrane, proceed the same way as for the removal of inhalation solution
FR
residues. As the cleaning agent, you can use a mixture of three parts of distilled water and
one part of bright vinegar.
C.
Cleaning the attached parts
IT
The attached parts (medicament holder, protective cap, mouthpiece and inhalation mask) are
then put into distilled water for 15 minutes. Afterwards, all parts are put onto a soft and clean
cloth and air-dried.
ES
NL
RU
PL
D.
Cleaning the main device
The main device is carefully wiped with a soft cloth. The contact pins can be carefully cleaned
using a cotton swab. Please ensure that the silicone seals are not damaged or removed.