Measurement References - FujiFilm SonoSite P16516-03 Manual Del Usuario

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 38
Imaging modes and exam types on transducers
Transducer
L25xp
L38xp
Importing and exporting
You must have administrative privileges to import or export labels, exam types, user accounts, and
system settings.
Measuring
Manual Trace and Auto Trace
The Manual Trace and Auto Trace measurements require that the Doppler Scale be set to cm/s in
Presets settings.
Measuring in Dual
You can measure across images in Dual only with a linear transducer.

Measurement references

Pulsatility Index (PI)
Zwiebel, W. J. Introduction to Vascular Ultrasonography, 6th Edition, W.B. Saunders Company, 2012.
PI = (PSV - MDV) / V (no units)
where:
PSV = peak systolic velocity
MDV = Minimum diastolic velocity
V = TAP (Time Averaged Peak) flow velocity throughout the cardiac cycle
About transducers
To remove a transducer
1 Pull the transducer latch up, and rotate it clockwise.
2 Pull the transducer connector away from the system.
4
Exam Type
Lung
Lung
Imaging Mode
Tabla de contenido
loading

Este manual también es adecuado para:

Sonosite x-porte

Tabla de contenido