Set output: Setting of universal output
(Out = SEr)
(Out = dAC)
(Out = dAC)
(Out = dAC)
Set correction: Adjustment of measurements
(°C / °F)
(%)
(hPa)
44
Universal output
oFF
Interface and analogue output off
(lowest possible power consump-
tion)
SEr
Serial interface activated
dAC
Analogue output activated
01.11 ... 91
Base address of the instrument
for serial interface communication
Conc
Analogue output relative to the
O
ppm
SAt
Analogue output relative to the
O
e.g. 1% O
Entry of the measurement value
2
at which the analogue output
must output 0V
e.g. 15% O
Entry of the measurement value
2
at which the analogue output
must output 1V
Zero point correction / offset of the
temperature measurement
oFF
No zero point correction
-5.0 ... +5.0
Zero point correction in °C
Slope correction of the voltage
measurement
oFF
No slope correction
-5.00 ...
Slope correction in %
5.00
Zero point correction / offset of the
temperature measurement
oFF
No zero point correction
-20 ... +20
Zero point correction in hPa
concentration in mg/litre or
2
saturation in % O
2
2
SD_315_DO_1 06/2020