10 Accessories
Electrodes
Miscellaneous
11 Technical data
11.1 Measuring properties
Measuring
Amperometric determination of dissolved oxygen (DO)
principle
Galvanic DO electrode (anode: platinum, cathode: lead)
Sensor
with integrated temperature sensor
Display range
Resolution
Accuracy
56
Description
Oxygen sensor, galvanic, Pt/Pb, cable length: 2 m
Oxygen sensor, galvanic, Pt/Pb, cable length: 10 m
Oxygen sensor, galvanic, Pt/Pb, cable length: 30 m
Description
Service set: 3 replacement membrane heads & KOH
electrolyte 100 ml
KOH electrolyte, 100 ml
PVC protective cap for depth measurement
Brass protective cap for depth measurement
USB data transmission cable
GSOFT 3050, Windows software (data logger /
data transmission)
AAA batteries, x 4
Demineralised water, 100 ml
Polypropylene measuring beaker, 100 ml
Calibration flask
O
concentration
2
O
saturation
2
Ambient pressure
Temperature
O
concentration
2
O
saturation
2
Ambient pressure
Temperature
O
concentration
2
Ambient pressure
Temperature
0 to 70 mg/litre or ppm
0 to 600%
0 to 1,200 hPa
-5 to +50°C
0.1 mg/litre or ppm
0.1%
1 hPa
0.1°C
a) 0 to 25 mg/litre: ± 1.5% of the measu-
rement value ± 0.2 mg/litre
b) 25 to 75 mg/litre: ± 2.5% of the measu-
rement value ± 0.3 mg/litre
3 hPa or 0.1% of the measurement value
(whichever is higher)
± 0.1°C
SD_315_DO_1 06/2020
Order No.
19805050
19805051
19805052
Order No.
724670
19801130
19805055
19805056
724620
724625
1950026
461275
384801
19805057