6.5

SETTING THE WEIGHING PARAMETERS

For proper use of the balance, it is important to make the correct weighing parameter settings
according to the operating environment.
This section describes the filter parameters, stability, auto-zero and weighing units.
In the settings menu, select the "Setup" function and then select "Weighing parameters".
© Adam Equipment 2018
In the "Weighing parameters" screen, the set
values (in the box to the right of the parameter)
are shown.
By touching the parameter key you want to
change, you can change the value.
 Filter: the filter function allows the user to
speed or slow down the balance's response
according to weighing requirements and
environmental
Level1, the response will be immediate, but
the balance will remain more sensitive to
environmental
ventilation and vibration. By increasing the
level, the response will be slower and the
indication more stable.
Filter 1: dosage conditions
Filter 2: stable conditions
Filter 3: unstable conditions
 Stability: this function allows you to adjust the
balance to the working environment. When
using
the
balance
environment, select level 0. The default level
is 2. Use level 3 for very disturbed
environments.
Stability 0: For stable environments.
Stability 1: For poorly stable
environments.
Stability 2: For unstable environments.
Stability 3: For heavily unstable
26
conditions.
By
disturbances
such
in
a
vibration-free
environments.
selecting
as