TABLE 1
GUIDANCE AND MANUFACTURER
The EX3 Stepper is intended for use in the electromagnetic environment specified below. The customer or the user of the EX3 Stepper should assure that it
is used in such an environment.
Emissions test
Compliance
RF  emissions
Group 1
CISPR 11
RF  emissions
Class A
CISPR 11
Harmonic
emissions
Not Applicable
IEC 61000-3-2
Voltage
fluctuations/
Not Applicable
flicker emissions
IEC 61000-3-3
TABLE 2
GUIDANCE AND MANUFACTURER
The EX3 Stepper is intended for use in the electromagnetic environment specified below. The customer or the user of the EX3 Stepper should assure that it
is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
'
-
S DECLARATION
ELECTROMAGNETIC EMISSIONS
Electromagnetic environment - guidance
The EX3 Stepper uses RF  energy only for its internal function. Therefore, its RF  emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
The EX3 Stepper is suitable for use in all establishments other than domestic and those directly
connected to the public low-voltage power supply network that supplies buildings used for
domestic purposes.
'
-
S DECLARATION
ELECTROMAGNETIC IMMUNITY
Compliance level
±6 kV contact
±4 kV contact
±8 kV air
±8 kV air
±2 kV for power supply
Not Applicable
lines
±1 kV for input/output
lines
±1 kV differential mode
Not Applicable
±2 kV common mode
<5% U
(>95% dip in U
)
Not Applicable
T
T
for 0,5 cycle
40%  U
(60% dip in U
)
T
T
for 5 cycles
70%  U
(30% dip in U
)
T
T
for 25 cycles
<5% U
(>95% dip in U
)
T
T
for 5 sec
3 A / m
3 A / m
Electromagnetic environment - guidance
Prior to operating the EX3 Stepper, touch the base of the stepper and
procedure table. Synthetic fabrics and carpets should be minimized.
Relative humidity should be at least 20%.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
If the user of the EX3 Stepper requires continued operation during
power mains interruptions, it is recommended that the EX3 Stepper
be powered from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of
a typical location in a typical commercial or hospital environment.