8. Configuration via display and operating unit
8.3 Configuring volume measurement
Requirement
EN
1. Open the operating menu with [↵].
Select "Basic setting" and confirm with [↵].
2. Select "Application" and confirm with [↵].
3. Select "Level" and confirm with [↵].
4. Select "Unit" and confirm with [↵].
5. Select length unit and confirm with [↵].
» Length unit is set.
6. Select "Density" and confirm with [↵].
7. Select "Density unit" and confirm with [↵].
8. Select density unit and confirm with [↵].
» Density unit is set.
9. Select "Density value" and confirm with [↵].
10. Set digit using [▲] [▼] and confirm with [↵].
» The cursor moves to the next digit.
» Repeat for each digit.
» Density value is set.
11. Go back two menu levels using [ESC].
Select "Mode" and confirm with [↵].
12. Select "Level" and confirm with [↵].
» Mode is set.
30
Length unit for the fill level is known
■
Medium density is known
■
Characteristic curve of the tank is known (→ See chapter
■
8.4 "Characteristic curves")
WIKA operating instructions, model DMSU21A