Adjustment Of Zero Point And Span - WIKA S-10 Manual De Instrucciones

Transmisor de presión
Ocultar thumbs Ver también para S-10:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

6. Adjustment of zero point and span

6. Adjustment of zero point and span
Only adjust the span-setting potentiometer if calibration equipment is available which has at least three
EN
times the accuracy of the pressure transmitter.
6.1 Preparation (figure A)
To gain access to the potentiometers, open the instrument as follows:
Disconnect the electrical connection
Remove the clamping nut
Carefully pull the instrument connector
Connect the instrument connector
(e.g. ammeter, voltmeter) according to the connection diagram.
6.2 Adjustment of zero point (figure B)
Go to the start of the measuring range.
Using potentiometer "Z", adjust the minimum output signal (e.g. 4 mA)
6.3 Setting the span (figure B)
Go to the end of the measuring range.
Using potentiometer "S", adjust the maximum output signal (e.g. 20 mA)
Check the zero point and if there is any deviation, re-adjust it.
Repeat the procedure until the zero point and the span are set correctly.
6.4 Finish the adjustment (figure A)
Disconnect the instrument connector
display unit.
Carefully push the instrument connector
damaging the wires or the seals. The seals must be clean and undamaged
in order to guarantee the given ingress protection.
Tighten the clamping nut
After the adjustment, check that the system is functioning correctly.
Recommended recalibration cycle: 1 yea
12
from the instrument.
.
from the instrument.
to the power supply and a display unit
from the power supply and the
onto the instrument, without
.
WIKA operating instructions pressure transmitter, model S-10
A
B

Hide quick links:

Tabla de contenido
loading

Tabla de contenido