Application Setup
The application can be customized for various user
preferences.
1.
Press the button corresponding to the
enter Configuration. (See Figure 74)
Figure 74
The Configuration Menu is now displayed. (See Figure 75)
Figure 75
2. Select the list item and press the button
corresponding to the
setting as desired.
3. To return to the Application home screen, press the
button corresponding to the
Sieve Configurations are defined below (defaults in bold).
ITEM
AVAILABLE SETTINGS
Accumulation %
Off, On
Retained
Accumulation
Off, On
Weight Retained
Fineness
Off, On
End Weight, Start
Calculate By
Weight
* If Start Weight is chosen, you must weigh original sample (or input
manually).
PAGE 24 OF 105
OPERATION CONTINUED
icon to
icon, to change the
icon.
COMMENTS
Print Accumulated %
Retained (on/off)
Print Accumulated %
Retained (on/off)
Print Fineness Modulus
(on/off)
Calculate Result with
End Weight or Start
Weight
Sieve Operation
To start sieve operation, follow the instructions below.
1.
Recall a sieve set from library to begin. (See Figure 76)
Figure 76
NOTE: A sieve set must be in library for the sieve
application to work. Each sieve to be used has
to be added to the set, even if they are not
used for fineness modulus calculation, as well
as the pan (with mm size 0). Sieve Weighing
always assumes that the procedure will be
from the grossest sieve (largest size) down to
the finest sieve (the pan). To create a sieve set,
see Library on page 26.
2. Press the button corresponding to the
start sieve weighing. (See FIgure 77)
Figure 77
3. Enter a sample ID if needed. (See Figure 78)
Figure 78
NOTE: The sample ID can be up to 30
alphanumeric characters and can be printed
on the output template.