Zero Check
Zero check is used when comparing subsequent
samples to an initial reference sample. In this case, the
Under Limit must be a negative value and the Over Limit
must be a positive value.
Place the reference item on the scale and press Tare.
Remove the reference sample and place the item to
be compared on the scale to determine if it is within the
Accept range.
Input/Output (I/O) Setup
The I/O's can be customized for various user preferences.
The I/O's are defined below (defaults in bold).
ITEM
Off, Zero, Tare, Clear Tare, Print, Unit,
Discrete Input 1
Accumulate
Off, Zero, Tare, Clear Tare, Print, Unit,
Discrete Input 2
Accumulate
Discrete Output 1 Off, Under, Over, Accept, Under/Over
Discrete Output
Off, Under, Over, Accept, Under/Over
2
Discrete Output
Off, Under, Over, Accept, Under/Over
3
Discrete Output
Off, Under, Over, Accept, Under/Over
4
NOTE: The I/O's will only work when the I/O
Option Board has been installed. The I/O option
board provides two isolated inputs and four
dry-contact normally open relay outputs which
can be used for simple process weighing.
Figure 28
1.
Press the
button to enter the Main Menu.
(See Figure 28)
2. With the button corresponding to the
down the list and highlight Application Mode. Enter
this submenu by pressing the button corresponding
to the
icon.
PAGE 13 OF 105
OPERATION CONTINUED
AVAILABLE SETTINGS
icon, go
3. In the Application Mode menu enter the Check sub
menu. (See Figure 29)
Figure 29
The Check submenu is now displayed. (See Figure 30)
Figure 30
4. Select the list item and press the button
corresponding to the
setting as desired.
FORMULATION
Use this application for compounding and recipe
making. The number of components can be 1 to 100.
Formulation has two available modes of operation:
Free Formulation and Recipe Formulation.
Press the
button until Formulation is displayed in
the upper left portion of the home screen.
Free Formulation (default)
This mode of Formulation allows the user to freely add
components. A recipe can also be saved and printed
when the formulation is finished.
THE FORMULATION HOME SCREEN
icon to change the
Main Display Line
Reference Fields