2-1. Precautions
To get the optimum performance from the balance and acquire accurate weighing data, note the following:
Install the balance in an environment where the temperature and humidity are not excessive. The
best operating temperature is about 20°C±2°C at about 45 ~ 60%RH relative humidity.
Install the balance where it is free of dust.
The weighing table should be solid (An anti-vibration table or stone table is ideal).
Install the balance on a horizontal table and check that it is not
tilted.
Install the balance in a stable place avoiding vibration and shock.
Corners of rooms on the first floor are best, as they are less prone
to vibration.
Install the balance where it is not affected by heaters or air
conditioners.
Install the balance where it is not exposed to direct sunlight.
Install the balance away from equipment which produces
magnetic fields.
Level the balance by adjusting the leveling feet and confirm it
using the bubble spirit level. Refer to the next page "How to adjust
the bubble spirit level".
Warm up the balance for at least 30 minutes. Plug in the AC
adapter as usual.
Calibrate the balance before use or after moving it to another
location. Refer to "8. CALIBRATION" in the instruction manual.
With the FZ-i WP / FX-i WP series
The dustproof and waterproof level of the balance is equivalent to IP65, and its second digit, "5",
corresponds to "having no harmful influence by receiving direct jet of water". Washing the balance with
strong water pressure or submerging it in water, may cause water to seep inside the balance and
cause a malfunction.
When installing and using the balance under conditions requiring dustproof and waterproof
performances, make sure that the AC adapter plug is fully inserted into the AC adapter jack and that the
terminal cover is attached to the RS-232C interface or the waterproof RS-232C cable
(AX-KO2737-500) is used.
If the RS-232C terminal cover is removed or the waterproof RS-232C cable (AX-KO2737-500) is not
used, protection against dust and water is not provided.
Caution: Do not install the balance where flammable or corrosive gas is present.
Leveling
foot
4
OK
Not OK
Bubble spirit level