Guidance and manufacturer's declaration – Electromagnetic Emissions
The HOSPI PLUS SUCTION ASPIRATOR is intended for use in the electromagnetic environment specified below. The
customers or the user of the HOSPI PLUS SUCTION ASPIRATOR should assure that it's used in such an environment.
Emissions Test
Irradiated / Conducted
emissions CISPR11
Irradiated / Conducted
emissions CISPR11
Harmonic emissions
EN 61000-3-2
Voltage fluctuations /
flicker emissions EN 61000-3-3
Guidance and manufacturer's declaration – Immunity Emissions
The HOSPI PLUS SUCTION ASPIRATOR is intended for use in the electromagnetic environment specified below. The
customers or the user of the HOSPI PLUS SUCTION ASPIRATOR should assure that it's used in such an envitonment.
Immunity Test
Level indicated by the
EN 60601-1-2
Electrostatic
+/-8kV contact
discharge
+/-15kV in air
(ESD) EN 61000-4-2
Electrical fast
+/-2kV power supply
transient / burst
lines
EN 61000-4-4
+/-1kV for input / output
lines
Surge
± 0,5kV
EN 61000-4-5
± 1,0kV differential mode
Volgate dips, short
<5% U
interruptions and
for 0,5 cycle
voltage variations on
40 % U
power supply input
for 5 cycle
lines EN 61000-4-11
70 % U
for 25 cycle
<5 % U
for 5 sec
Magnetic field
30A/m
(50/60 Hz)
EN 61000-4-8
Note UT is the value of the power supply voltage.
Compliance
Electromagnetic environment - guidance
Group 1
The HOSPI PLUS SUCTION ASPIRATOR only used RF energy
only for its internal functioning. Therefore its RF emissions are very
low and are not likely to cause any interference in nearby electronic
equipment.
Class [B]
The appliance is suitable fopr use in alla establishments included
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
Class [A]
domestic purposes.
Complies
Compliance Level
The
device
change its state
The
device
change its state
The
device
change its state
(>95% dip U
)
- -
T
T
(60% dip U
)
T
T
(30% dip UT )
T
(>95% dip U
)
T
T
The
device
change its state
ENGLISH
Electromagnetic environments - guid-
ance
doesn't
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
doesn't
Mains power quality should be that of
a typical commercial or hospital enviro
ment.
doesn't
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Mains power quality should be that of a
typical commercial or hospital environ-
ment. If the user of the HOSPI PLUS
SUCTION ASPIRATOR requires contin-
ued operation during power mains inter-
ruptions, it is recommended that the de-
vice be powered from an uninterruptible
power supply or a battery.
doesn't
The power frequency magnetic field
should be measured in the intended in-
stallation location to assure that it's suf-
ficiently low.