CLEANING THE MAIN UNIT

To clean the device external parts always use a cotton cloth dampened with detergent. Don't use abrasive or
solvent detergents.
ATTENTION: During cleaning make sure that liquids do not come into contact with the membrane keyboard
(only in versions with footswitch and flux deviator) and adjacent areas as this would damage the component,
with possible infiltration of the liquid inside the device..
The symbol
positioned in the casing near the membrane keyboard requires the reading of the user
instructions before each use.
PARTICULAR CARE SHOULD BE TAKEN TO ENSURE THAT THE INTERNAL PARTS OF THE
EQUIPMENT DO NOT GET IN TOUCH WITH LIQUIDS. NEVER USE LIQUIDS (e.g. DETERGENTS AND/
OR SANITISING SUBSTANCES) TO CLEAN THE MAIN UNIT (ESPECIALLY NEAR THE MEMBRANE
KEYBOARD) AS THEY MAY PENETRATE INSIDE THE DEVICE
During all clearing operations use protection gloves and apron (if need be, also wear a face mask and glasses)
to avoid getting in contact with contaminating substances (after each utilization cycle of the machine).

ACCESSORIES SUPPLIED

N°2 COMPLETE ASPIRATION JAR
CONICAL FITTING
SILICON SET TUBES 8 mm x 14 mm
ANTIBACTERIAL AND HYDROFOBIC FILTER
SAFETY TRAP
FOOTSWITCH CONTROL (for versions equipped with footswitch)
EUROPEAN POWER SUPPLY CORD (H05VV-F - 2x0.75mm² - 2mt)
REPLACING THE ANTIBACTERIAL FILTER:
passage of liquids into the same filter. If you suspect the filter may have been contaminated and/or got wet
or discoloured, always remove and replace the filter. If the equipment is to be used on patients with unknown
pathological conditions or should you evaluate the possibility of indirect contamination, remove and replace
the filter after each utilization. The filter is not designed for decontamination, disassembly and/or sterilization. If
you suspect the filter may have been contaminated and/or got wet or discoloured, always remove and replace
the filter. If the equipment is to be used on patients whose pathologies are known and not implying any indirect
contamination risks, we recommend to remove and replace the filter at the end of each work shift or else every
month, even if the equipment has not been used. 4000ml or 5000ml complete jar versions are available on
request. Versions fitted with FLOVAC® 2000ml or 3000ml disposable collection systems (including a re-usable
rigid polycarbonate container and a disposable Liner) are also available on request.
SAFETY TRAP with 220 ml capacity to collect the liquid that could leak from the overflow valve of the vessel.
This ensures additional protection of the filter and pump. The trap is completely removable and autoclavable.
Not available in versions equipped with FLOVAC disposable collection system.
WARNING: Suction tubes for insertion in the human body purchased separately from the machine should
comply with ISO10993-1 standards on material biocompatibility.
WARNING: The medical device is provided without a specific suction probe. If this device must be used with a
specific suction probe, the end user is responsible for making sure it complies with the EN 10079-1 regulation.
Aspiration jar: The mechanical resistance of the component is guaranteed up to 30 cycles of cleaning and
sterilization. Beyond this limit, the physical-chemical characteristics of the plastic material may show signs of
decay. Therefore, we recommend that you to change it.
Silicone tubes: the number of cycles of sterilization and/or cleaning is strictly linked to the employment of the
said tube. Therefore, after each cleaning cycle, it is up to the final user to verify whether the tube is suitable
for reuse. The component must be replaced if there are visible signs of decay of the material constituting the
said component.
ENGLISH
The filter is made of hydrophobic material that stops the