Dräger Polytron 7000 Instrucciones De Uso página 84

Transmisor para sensores electroquímicos
Ocultar thumbs Ver también para Polytron 7000:
Tabla de contenido
Polytron 7000 funcionamiento con PROFIBUS PA
PROFIBUS PA:
Coding of the Physical Block Parameter DIAGNOSIS
Octet
Bit
Mnemonic
1
0
DIA_HW_ELECTR
1
1
DIA_HW_MECH
1
2
DIA_TEMP_MOTOR
1
3
DIA_TEMP_ELECTR
1
4
DIA_MEM_CHKSUM
1
5
DIA_MEASUREMENT
1
6
DIA_NOT_INIT
1
7
DIA_INIT_ERR
2
0
DIA_ZERO_ERR
2
1
DIA_SUPPLY
2
2
DIA_CONF_INVAL
2
3
DIA_WARMSTART
2
4
DIA_COLDSTART
2
5
DIA_MAINTAINANCE
2
6
DIA_CHARACT
2
7
IDENT_NUMBER_VIOLATION Set to 1 (one), if the Ident_Number of the running cyclic data transfer and the
3
0...7 reserved
4
0...6 reserved
4
7
EXTENSION_AVAILABLE
Coding of the Physical Block Parameter DIAGNOSIS_EXTENSION
Octet
Bit
Mnemonic
1
0
DIA_SENSOR_ERR
1
1
DIA_OTHER
1
2
DIA_SOFTWARE_ERR
1
3
DIA_CALIBRATION_ERR
1
4
DIA_MAINTAINCANCE_SOON Maintenance required soos
84
Description
Hardware failure of the electronic
Hardware failure mechanics
Motor- temperature too high
Electronic temperature too high
Memory error
Failure in measurement
Device not initialized (No selfcalibration)
Selfcalibration failed
Zero point error (limit position)
Power supply failed (electrical, pneumatic)
Configuration not valid
New-start-up (warmstart up) carried out.
Re-start-up (coldstart up) carried out.
Maintenance required
Characterization invalid
value of Physical Block IDENT_NUMBER_SELECTOR parameter are different.
Reserved
Reserved
More diagnosis information is available
Description
Sensor failure
An error has occurred that was not identified.
The software has detected an error.
An error occurred during calibration of the device or a calibration error has
been detected during operation of the device.