Lovibond Tintometer MD 100 Instrucciones página 41

Ocultar thumbs Ver también para Tintometer MD 100:
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 94
User calibration
GB
Sample
Zero
Test
METHOD
METHOD
0.0.0
CAL
Zero
Test
METHOD
METHOD
RESULT
RESULT
CAL
On
Off
Mode
Zero
Test
CAL
RESULT
RESULT + x
+ x
On
Off
MD1X0/200_1b 9b/2020
Place the vial in the sample chamber.
Pay attention to the positioning.
press
flashes for approx. 8 seconds
The display shows the following in alternating mode:
Perform the measurement with a known standard concentration.
press
flashes for approx. 3 seconds
The result is shown in the display, alternating with CAL.
If the result matches the value of the standard used (within the relevant
tolerance), exit calibration mode.
press
If the result is outside of the value (taking into account the tolerance),
change the displayed value:
Press 1 x increases the displayed value by 1 digit.
Press 1 x decreases the displayed value by 1 digit.
Press the corresponding key until the reading equals the value of the
calibration standard.
The result is shown in the display, alternating with CAL.
press for more than 4 seconds
The user calibration is calculated and stored.
41
loading

Este manual también es adecuado para:

Tintometer md 110Tintometer md 200