Functional Description; Operation; Scroll Memory; Performing Zero - Lovibond Tintometer MD 100 Instrucciones

Ocultar thumbs Ver también para Tintometer MD 100:
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 94

Functional description

GB
On
Off
Mode
Sample
MD1X0/200_1b 9b/2020

Operation

Switch the unit on
Selecting a method
Scroll Memory (SM)
To avoid unnecessary scrolling for the required test method, the instru-
ment memorizes the last method used before being switched off. When
the instrument is switched on again, the scroll list comes up with the
last used test method first.
Note:
The description of the methods and reagent handling is not included
in this manual; it is contained in a separate document.
Fluoride method
Caution:
A calibration of the method has to be performed before the first use
4
(see
calibration mode Fluoride). The same batch of SPADNS reagent
solution must be used for adjustment and test. The adjustment process
needs to be performed for each new batch of SPADNS reagent solution.

Performing zero

Fill 24 mm vial with 10 ml sample
Close vial
Place sample vial in the sample chamber.
Pay attention to the positioning.
33
loading

Este manual también es adecuado para:

Tintometer md 110Tintometer md 200