Figure 5.
Filter life can be displayed in both hours and minutes or as a percentage using the Settings
button located in the lower left section of the touch screen. See Figures 6 and 7.
Figure 6.
Figure 7.
When the maximum filter life is consumed, the following warning screen will appear indicating
that a filter replacement is needed. After inserting a new VisiClear™ Filter, press the green
check mark to clear the warning and resume normal operation. During operation, if a filter
expires the system will not turn off but will remain running until the system is powered down
or 6 hours passes, whichever comes first. At that point, the VisiClear™ will no longer operate
until a new filter is inserted into the system. It is always best practice to replace an expired
filter immediately; however, a validated contingency life is built into the filter to ensure no
interruption to the operative procedure. See Figure 8.