Cleaning And Disinfection; Sterilising The Hand Grip - Dräger Polaris 50 Manual De Usuario

Ocultar thumbs Ver también para Polaris 50:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 80
Polaris 50 Mobile
DANGER
Risk of death from electric shock.
Do not plug in any damaged power cables.
If the power cable shows damage, immediately replace
it with a new one.
The connection voltage and frequency must match the
data on the type plate.
Only connect to the power grid with protective earth
conductor (PE)
CAUTION
When changing locations
Put the terminal into its lowest position.
Unlock the casters.
Be careful not to run over anything, including the
connection cable.
Hold firmly during transport within the clinic.
Be particularly careful about slopes,sills, bumps and
other obstacles.
CAUTION
When it is not in use, wind the power cable onto the
cable holder.
Insert the power cable.
Connect cable to the power grid.
8.

CLEANING AND DISINFECTION

DANGER
Risk to life from electric shock
Before disinfection and cleaning, switch off the power
connection and ensure it cannot be activated
unintentionally.
CAUTION
Material damage due to incorrect cleaning
For cleaning, only use products that do not impair the
operation of the luminaire.
For cleaning, do not use solvent-based, chlorine-
based or abrasive detergents, because they can crack
plastic parts and cause other damage.
The cleaning agents must be approved for use on
plastics such as PC, PMMA, PA and ABS.
Concentrated disinfectant can damage the luminaire.
For concentration and application times, check the
information provided with the product used.
The wrong cloth may cause scratches.
RECOMMENDED DISINFECTANTS
Dismozon Plus
Kohrsolin Extra
Lysoformin
Microbac Tissues
Mikrozid Sensitive Liquid
CAUTION
Dirt reduces the light strength
Keep the cover clear through regular cleaning.
Only cleaning by wiping is allowed.
Clean the clear PMMA cover with a chamois soaked in
windscreen wiper liquid.
CAUTION
To minimise the risk of disease transmission, applicable
health and safety regulations and the requirements of the
national hygiene and disinfection authorities must be
observed in addition to these instructions.
8.1

Sterilising the hand grip

Sterilisation must be carried out according to ISO 17665-1
(Sterilisation of health care products in moist heat).
ENG
35
Tabla de contenido
loading

Tabla de contenido